Nonin Sensors - 8000J

Description

Nonin 8000J reusable pulse oximetry sensor system combines a flexible silicone sensor with single-use FlexiWrap adhesive to deliver secure, accurate SpO₂ and pulse rate monitoring for neonate to adult patients.

Size:
8000J (Adult), 8008J (Infant), 8001J (Neonatal)
Category:
Oximetry & monitoring
Respiratory therapy
Sleep diagnostics
8000J

Reusable Pulse Oximetry Sensor System

The Nonin 8000J Reusable Pulse Oximetry Sensor System delivers the ease of a disposable sensor combined with the long-term value of a reusable design. The flexible silicone sensor is paired with a single-use FlexiWrap adhesive, providing a secure and comfortable fit during extended monitoring periods, including sleep studies. The 8000J Series is available in configurations sized for neonates through adults to ensure proper placement and consistent readings across patient groups.

System Configurations
Adult Flex System (8000J Sensor + 8000JFW FlexiWrap)

- Cable Length: 1 m / 3 m / WristOx2

- Patient Range: > 20 kg (44 lbs)

- Preferred Application: Index, middle, or ring finger

Infant Flex System (8008J Sensor + 8008JFW FlexiWrap)

- Cable Length: 1 m

- Patient Range: 2 to 20 kg (4.4 to 44 lbs)

- Preferred Application: Toe

Neonatal Flex System (8001J Sensor + 8001JFW FlexiWrap)

- Cable Length: 1 m

- Patient Range: < 2 kg (4.4 lbs)

- Preferred Application: Lateral aspect of the foot

Accuracy Specifications
SpO₂ Accuracy (Adults/Pediatrics)

Range        Oxygen Saturation (Arms*)

70–100%        ±3

70–80%        ±3

80–90%        ±2

90–100%        ±2

Low Perfusion Accuracy

- SpO₂: 70–100% ±2 digits (Arms*)

- Pulse Rate: 40–240 BPM ±3 digits (Arms*)

- Pulse Rate Low Perfusion: 40–240 BPM ±3 digits (Arms*)

*±1 Arms represents 68% of the population at zero bias.

- Accuracy testing was performed under no-motion conditions and is based on Nonin’s PureSAT® SpO₂ technology and PureLight® sensor technology.

- Additional accuracy details are available in the sensor accuracy document (operator’s manual CD).

- Combined sensor/oximeter specifications should be referenced in the applicable oximetry system manual.

Environmental Specifications
Temperature

- Operating: –20 °C to 50 °C (–4 °F to 122 °F)

- Storage/Transport: –40 °C to 70 °C (–40 °F to 158 °F)

Humidity

- Operating & Storage/Transport: 10% to 95% non-condensing


Measurement Wavelengths & Output Power

- Red: 660 nm @ 3 mW nominal

- Infrared: 910 nm @ 3 mW nominal

This information is especially useful for clinicians performing photodynamic therapy.
Compliance

- Complies with ISO 10993-1

- Not made from natural rubber latex

Standard delivery:

Orders are typically processed within 1-2 business days and delivered within 3-5 business days. Delivery may be delayed during weekends or public holidays.

Shipping charges:
  • Orders from 0$ to 149$: shipping fee 10$.
  • Orders 150$ and above. Free shipping.
International shipping:

Currently, we do not offer international shipping.

Shipping issue and support:

If you experience any delays, damage, or missing items, please contact us immediately at +6593633572. We're here to ensure your shopping experience remains smooth and satisfactory.

Similar Products

Explore other devices and masks that may suit different clinical scenarios, patient preferences or care settings.